※Product Category:400Series >> EC-400W Electrochemical oxygen analyzer Source:www.tgking.net |
|
新建网页 2
|
|
The EC-400
Electrochemical Micro Oxygen Analyzer is an online analyzer that uses a
new type |
of intelligent detection
instrument that combines imported sensors with a single chip computer,
|
with ppm display, and is widely
used for the detection of oxygen content in industrial production |
processes such as air
separation and nitrogen production, fertilizer, petrochemical industry,
and |
biological fermentation. The
fuel pool oxygen sensor is composed of highly active oxygen |
electrodes and lead electrodes,
which are immersed in an electrolyte solution. Using this method |
to measure oxygen can be
unaffected by the reducing gases in the measured gas. The EC-400 |
electrochemical micro oxygen
analyzer eliminates many sample gas processing systems. |
◆
The instrument adopts a new imported fuel cell oxygen sensor,Fast
measurement speed, simple user calibration and debugging |
◆
The instrument adopts a large screen LCD display screen with full Chinese
operation menu |
◆ The sensor is not
interfered by impurities such as H2, CO, H2S, NOX, etc |
◆
Real time clock display, with 64 paperless data records and storage in
the background |
◆
The output current is 4-20mA, which can be connected to a PLC,
Available standard gas for online calibration |
◆
Upper and lower limit alarm points can be set, with alarm contact
output |
|
Measurement range :
0~1000ppm O2 (three digit LCD display in Chinese, maximum overload to
2000ppm, imported sensor) |
Basic error : ± 2.0%
FS |
Repeatability : ± 2.0% FS |
Response time : from 100
PPM → 1 ppm<240 seconds |
Sample gas flow control
: 250-300ml/min |
Sensor life :>2 years under normal use |
Power supply : AC
220V+10%; 50Hz |
Environmental conditions :
0-50 ℃; Relative humidity<90% |
Shell protection : metal shell sprayed with plastic |
Alarm setting : Any setting
within 0~1000ppm |
Power consumption : less
than 15W, fuse 3A |
Overall dimension : 145 ×
145 × 250 (wide × high × Depth) mm |
Opening size : 138 × 138
(wide × Height) mm |
Output signal : 4-20mA, gear
I: 0-10ppm, gear II: 0-100ppm, gear III: 0-1000ppm |
Ambient temperature
compensation correction : AD590 temperature module, computer automatic
correction |
|
The following
gases have no effect on the measurement results |
HCCH |
Air |
NH3 |
Ar |
C2H8 |
CO |
H2 |
He |
N2 |
C3H6 |
C2H4 |
SF6 |
ASH3 |
PH3 |
SIH4 |
HC |
CH2O |
|
|
|
|
|
|
|
The following gases have no interference with the sensor signal
output, but exposure to the above recommended values (regardless
of time and concentration) can reduce the sensor life by 3-4
months or even fail |
C4H6 |
CO2 >1000ppm |
HCL>1000ppm |
HCN>1000ppm |
HF>1000ppm |
H2S>10ppm |
SO2>10ppm |
|
The
following gases have a small impact on the sensor life, but
cross interference with the oxygen sensor results in a reading
deviation (1 ppm causing 2 ppm error) |
CL2 |
NO2 |
|
|
|
|
|
|
|
|
|